PN T900-013, Rev A
Chassis
Specication Description
Dimensions 18 cm (7") H x 43 cm (16.8") W x 67 cm (26.5") D
Weight 15.9 kg (35 lb)
Power 100–120 VAC, 50/60 Hz
200–240 VAC, 50/60 Hz
Outputs
Specication Description
Display LCD graphics, backlit, 100 mm x 150 mm
Printer Port Concentration, chromatogram, and diagnostic reports
Serial PLC Port Concentration data and general alarm via RS-232
4–20 mA (optional) Concentration signals
ta7000R PLC Output Format
The concentration data, status, and other detail information is sent from Port2
of the ta7000R automatically at the end of every run. The format used is called
Comma Separated Variable (CSV), a common MS Excel format.
The elds within each record are dened below. Fields 1–6 are the record iden-
tication elds and Fields 7–8 are the impurity elds.
The next impurities will be at Fields 9–10, 11–12, 13–14, 15–16, and 17–18. This
string is terminated by a carriage return (CR), and line feed (LF) characters.
The parameters are 9,600 Baud, 8 Data Bits, 1 Stop Bit, and No Parity.
Field 1 Date & Time of RUN
Field 2 Analyzer Serial #
Field 3 Run/Maintenance
Field 4 EDL/Single Analysis
Field 5 N/A
Field 6 ta7000 Error Status
Field 7 Impurity Name
Field 8 Concentration in PPB x 100
Specications | 5-3