EasyManuals Logo

Ametek ta7000R User Manual

Ametek ta7000R
57 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
PN T900-013, Rev A
General Safety Considerations
Components and Functions
The following safety items must be observed to avoid personal injury and to prevent
damage to the ta7000R Monitor. All personnel involved in installation, operation or
service of the ta7000R should be thoroughly familiar with these items. These consider-
ations are in addition to all site specic safety procedures.
Mercury Vapor
In RGD-equipped instruments, be sure the Mercury vapor scrubber is con-
nected before operating the analyzer and plan on replacing the scrubber after
6 months of continuous operation or with each HgO bed replacement.
UV Light
In RGD-equipped instruments, do not look directly at the ultraviolet lamp.
Since UV light causes eye damage, use appropriate safety glasses whenever
the UV lamp is exposed.
Electrical Hazard
Be sure that the top cover is securely in place before operating the analyzer.
Since some service procedures require the cover to be removed while the
power is on, be aware that electric shock can occur. Therefore, only trained
or otherwise qualied personnel should service the analyzer. The analyzer is
supplied with a 3-conductor power cord providing a protective earth ground
connection when plugged into a properly wired receptacle. Proper receptacle
grounding must be veried.
Contents | ix

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek ta7000R and is the answer not in the manual?

Ametek ta7000R Specifications

General IconGeneral
BrandAmetek
Modelta7000R
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals