PN T900-013, Rev A
Each eld is dened as:
DDDD-DD-DD TT:TT:TT,#,R,S,A,EEEEEEEE,NNNNN,C,
DDDD-DD-DD = Date (2000-03-23)
TT:TT:TT = Time (12:23:44)
# = Analyzer Serial Number
R = Run or Maintenance (Calibration) Mode
A = EDL or Single Analysis Mode
S = N/A
EEEEEEEE = Eight digits of error status (0 = Not Error, 1 = Error)
1. General Alarm (MSD)
2. Spare
3. Spare
4. Spare
5. Spare
6. Spare
7. Spare
8. Spare (LSD)
NNNNN = Name of impurity
C = Concentration in PPB x 100
An example of this output string is:
2001-08-16 14:54:13,54,M,E,2,00000000,CH4 ,441,CO2 ,5332,NMHC,172
Communication Ports
The ta7000R has three (3) communication ports (COM 1, COM 2, COM 3), which
are 9-pin male D-type connectors. Refer to Chapter 2 for more information.
Parallel Printer Port (LPT1)
The Parallel Printer Port uses a standard 25-pin female D-type connector. The
printer port follows standard parallel printer protocols. Refer to Chapter 2 for
more information about printer requirements.
5-4 | ta7000R Gas Purity Monitor