EasyManua.ls Logo

Anritsu MS2665C - 7.2 Parts, PC board installation

Anritsu MS2665C
178 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-3
7.2 Parts, PC board installation
7.2.1 Cabinet disassembly/assembly
Fig. 2-3-1 and 2-3-2 (for MS2665C) shows mechanical structure of the instruments (For MS2667/68C, see Fig. 3-3-1 and
3-3-2). To disassemble cabinet, perform procedures below.
(1) Remove eight feet.
(2) Draw enclosing cabinet backward and remove it.
(3) Remove rear panel
To assemble, perform inversely.
7.2.2 Parts/PCBs fitting for each option
7.2.2.1 Option 04 Hi-speed time domain sweep
Related units :
(1) A05 SCAN/AD
Additional parts :
(1) A0501 HI-SPEED AD constituent list (34Y106688)
Procedure :
Refer to paragraph 2.3.5 (for MS2665C) or paragraph 3.3.5 (for MS2667/68C).
(1) Connect A0501-X1 and A0501-X2 to A05-X6 and A05-X7.
(2) Fix A0501 to A05 with the strut and four screws.
(3) Turn on power, change option setting in Maintenance menu to OPT04 ON.
7.2 Parts, PC board installation