Section 2 MS2665C
2-24
2.2.5.3 IF1 (ATT), IF2 (AMP) adjustment
Required equipment :
(1) 6769B swept frequency synthesizer
(2) ML4803A Power meter
(3) MA4705A Power sensor
(4) IBM PC/AT compatible
Setup :
(1) Connect the spectrum analyzer RF Input to 6769B RF OUTPUT
(2) Connect the spectrum analyzer and the personal computer with RS-232C interface
Procedure :
(1) Turn the spectrum analyzer on, while pushing 0 key.
(2) Initialize the spectrum analyzer :
1) Enter Preset menu with Preset key,
2) Initialize the spectrum analyzer completely with F1 key.
(3) Calibrate the spectrum analyzer using its internal calibration function :
1) Enter Cal menu with Shift key and 0 key,
2) Calibrate the spectrum analyzer by pushing F1 key.
[ Mixer harmonic order 1 ]
(4) Set mixing mode to Band 1- :
1) Enter Frequency menu by pressing Frequency key, and open its second page with More key,
2) Enter Band menu with F1 key,
3) Set mixing mode to Band 1- by pushing F3 key.
(5) Set the spectrum analyzer to :
Center frequency, 5.00 GHz
Span, 1 MHz
Set the 6769B output frequency to 5.00 GHz (CW).
(6) Adjust the 6769B output level to make power meter reading -10 dBm at the end of cable feeding the signal to the
spectrum analyzer, and then connect the cable to the spectrum analyzers RF Input.
(7) Tune the spectrum analyzers pre-selector, using its pre-selector auto tune function :
1) Enter Cal menu with Shift + 0 key,
2) Enter Pre-selector Tuning menu with F6 key,
3) Tune the pre-selector by pushing F1 key.
(8) Read marker level using peak search function (Press Peak Search key), and if maker reading is within -10 dBm ±0.2
dB, skip next procedure.