Section 7 Options
7-6
7.2.2.4 Option10 Centronics interface
Related units :
(1) A03 CPU
Additional parts :
(1) A04 PMC/CENTRONICS constituent list (34Y106692A or 34Y106692B), refer to paragraph 2.2.1.3 (For MS2665C)
or paragraph 3.2.1.3 (For MS2667C) or paragraph 4.2.1.3 (For MS2668C).
Procedure :
Refer to paragraph 2.3.2.
(1) Connect A04-X1 and A04-X2 to A03-X2 and A03-X3.
(2) Fix A04 to A03 and the front frame with strut and five screws.
(3) Remove a cover on the back panel, attach the PARALLEL connector of A04-W1 cable to the back panel with both a
patch and the CENTRONICS plate (four screws).
(4) Connect A04-W1 cable to A04-X4.
(5) Do not remove PMC cap on the front panel.
(6) Turn on power, change the option setting in the Maintenance menu to OPT10 ON.