3-31
(9) Set IF Gain-1 and IF Gain-2 values to 0.
1) Press F4 (IF Gain-1) + 0 + enter keys and on display appears a writing IF Gain1 set to 0.
2) Press F5 (IF Gain-2) + 0 + enter keys and on display appears a writing IF Gain2 set to 0.
(10) Adjust IF Gain 1 and IF Gain 2 so that marker reading becomes -10 dBm ±0.5 dB.
If the level is lesser than this, the level can be raised by increasing the Number of IF Gain-2 (F5) from 0 to 255 in
single whole numbers.
Press F5 + number (0 to 255) + enter keys.
If the level is greater than this, the level can be lowered by increasing the Number of IF Gain-1 (F4) from 0 to 255
in single whole numbers.
Press F4 + number (0 to 255) + enter keys.
[ Writing the compensation values of IF Gain1 and IF Gain2 to Flash Memory ]
(11) After the above adjustment is done, Press F6 (return) + F1 (Version & options) keys.
(12) Press Key more 3 times. MAINTENANCE (4/4) page appears.
(13) Press the cursor down key (F5) till it falls on Save model and Option Data? and after that press F1 (Select) Key.
(14) On pressing the above key F1 key turns to SAVE. Press again and F1 key turns Really save? at this stage press
F2 (Yes).
(15) The display shows a message Now saving, Wait ..........
(16) Wait till this message disappears and after that Switch OFF the power supply of the spectrum analyzer.
(17) Now once again Switch ON the power supply of the spectrum analyzer with Preset key depressed.
3.2 Troubleshooting