CAUTION:The selected memory location will be over-
written by the SAVE SETUP operation. No warning is
given.
START
CAL
Opens a calibration menu. Use the Up/Down Arrow
key and ENTER key to select either COAX or
WAVEGUIDE calibration. Then select the desired
setup configuration from the next menu screen.
For Coax Calibration, follow the text in the message
area that instructs you to do the following:
q
Connect OPEN, Press ENTER.
The Site Master will measure the calibration
“open” that you must attach to the end of the test
port .
q Connect SHORT, Press ENTER
The Site Master will measure the calibration
“short” that you must attach to the end of the test
port.
q Connect LOAD, Press ENTER
The Site Master then measures the 50W termina-
tion (load) that you must attach to the end of the
test port.
For Waveguide Calibration, follow the text in the
message area that instructs you to do the following:
q
Connect SHORT 1, Press ENTER
The Site Master will measure the calibration
“short” that you must attach to the end of the test
port.
q
Connect SHORT 2, Press ENTER
The Site Master will measure the calibration
“short” that you must attach to the end of the test
port.