Quick Start Guide 2-16System Settings
MS20xxC UG PN: 10580-00305 Rev. J 2-33
MS20xxC Master Reset Instructions
1. Turn the MS202xC or MS203xC On.
2. Press the Shift key then the System (8) key.
3. Press the System Options soft key.
4. Press the Reset soft key.
5. Press the Master Reset soft key.
6. A dialog box is displayed on the instrument screen to warn that all settings will be
returned to factory default values, and that all user files will be deleted.
7. Press the Enter key to complete the master reset, or press the Esc key to abort.
8. After several seconds (which can grow to several minutes if a very large number of
measurements have been saved in the instrument), the instrument reboots.
2-16 System Settings
To access the System menu, press the Shift key, then the System (8) key. Refer to
Chapter 6, “System Operations ”.
Power On Self Test
At power on, the VNA Master runs through a series of quick checks to ensure that the system
hardware is functioning properly. Refer to “Power On Self Test” on page 6-13 and to “Self
Test” on page 6-4.
Vector Network Analyzer Mode
The “Application Options Menu (VNA Mode)” on page 6-7 contains the selections for Units,
External Reference, Trace Label, Measurement Gain Range, and Time Domain (if applicable).
Time Domain
When the Time Domain Option (Option 2) is enabled, certain aspects of the measurements
are determined by the settings in this menu. Refer to the Vector Network Analyzer
Measurement Guide (listed in Appendix A) for a detailed description of this menu.
Spectrum Analyzer Mode
The “Application Options Menu (SPA Mode)” on page 6-9 contains the Impedance setting
selection and an Auto Ref Level selection.
Frequency Blanking
This feature is available with Option 7 installed. Refer to “Frequency Blanking” on page 4-5.
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu VNA master and is the answer not in the manual?

Anritsu VNA master Specifications

General IconGeneral
BrandAnritsu
ModelVNA master
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals