4-3 Calibration Setup Secure Data Option 7
4-2 PN: 10580-00305 Rev. J MS20xxC UG
4-3 Calibration Setup
After any user calibration, the VNA Master automatically writes the calibration file to
internal memory. This is done so that when the instrument is turned Off and then back On,
the user calibration data are recalled, and the calibration can be applied.
With Option 7 enabled, however, the VNA Master cannot write to internal memory.
Therefore, the user calibration file is not automatically saved. With Option 7 enabled, when
the instrument is turned Off and then back On, the user calibration data are not recalled, and
the calibration cannot be applied. To save and recall the calibration, use the file save menu to
save the setup with calibration to USB memory. Then use the file recall menu to retrieve the
calibration.