2-12 VNA Master Connectors Quick Start Guide
2-26 PN: 10580-00305 Rev. J MS20xxC UG
Port 1 (50 ohm)
This Type-N connector (refer to Figure 2-13, item “6” on page 2-20) provides the input/output
50 ohm interface for transmission and reflection measurements of the Vector Network
Analyzer at Port 1. Maximum input is +23 dBm at ±50 VDC. Bias Tee output is also
available from this port (with Option 0010). The Port 1 green LED (refer to Figure 2-13,
item “17” on page 2-20) indicates (is illuminated) when the port is transmitting power.
To prevent damage to your instrument, do not use pliers or a plain wrench to tighten the
Type-N connector. Do not overtighten the connector. The recommended torque is 12 lbf·in to
15 lbf·in (1.36 N·m to 1.70 N·m).
Port 2 (50 ohm)
This Type-N connector (refer to Figure 2-13, item “9” on page 2-20) provides the input/output
50 ohm interface for transmission and reflection measurements of the Vector Network
Analyzer at Port 2. Maximum input is +23 dBm at ±50 VDC. Bias Tee output is also
available from this port (with Option 0010). The Port 2 green LED (refer to Figure 2-13,
item “10” on page 2-20) indicates (is illuminated) when the port is transmitting power.
To prevent damage to your instrument, do not use pliers or a plain wrench to tighten the
Type-N connector. Do not overtighten the connector. The recommended torque is 12 lbf·in to
15 lbf·in (1.36 N·m to 1.70 N·m).
Type K Connectors (Option 0011) for VNA Ports
To prevent damage to your instrument, do not use pliers or a plain wrench to tighten the
Type-K connector. Do not overtighten the connector. The recommended torque is 8 lbf·in
(0.9 N·m or 90 N·cm).
Bias Input Port 1
The BNC female connector (refer to Figure 2-13, item “7” on page 2-20) is used for external
bias tee input that will be routed to Port 1 of the Vector Network Analyzer. Maximum input is
±50 VDC and 500 mA.
To prevent damage to your instrument, do not use pliers or a wrench to tighten the BNC
connector. Do not overtighten the connector.
Bias Input Port 2
The BNC female connector (refer to Figure 2-13, item “8” on page 2-20) is used for external
bias tee input that will be routed to Port 2 of the Vector Network Analyzer. Maximum input is
±50 VDC and 500 mA.
To prevent damage to your instrument, do not use pliers or a wrench to tighten the BNC
connector. Do not overtighten the connector.
Bias Status LED
This LED (refer to Figure 2-13, item “16” on page 2-20) illuminates green when Internal or
External Bias is selected. It illuminates red for any overload condition (current or voltage).
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu VNA master and is the answer not in the manual?

Anritsu VNA master Specifications

General IconGeneral
BrandAnritsu
ModelVNA master
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals