Patient monitor Users manual
50
Alarm messages
Cause
Level
Temp labeloff
Temp label sensor may be
disconnected from user or monitor
Low
Temp label over
measuring range
Temp labelover measuring range
Temp label
below
measuring range
Temp label
below measuring range
Temp self checking error
Temp calibration failure
Temp communication
error
Temp measurement error or
communication error
Medium
NIBP
Physiological alarm:
Alarm messages
Cause
Level
NIBP Sys high
NIBP Sys measuring value is above
high alarm limit
Medium
User-Selectable
NIBP Sys low
NIBP Sys measuring value is below
low alarm limit
NIBP Dia high
NIBP Dia measuring value is above
high alarm limit
NIBP Dia low
NIBP Dia measuring value is below
low alarm limit
Medium
User-Selectable
NIBP Map high
NIBP Map measuring value is above
high alarm limit
NIBP Map low
NIBP Map measuring value is below
low alarm limit
Technical alarm 1displayed in the message area:
Alarm messages
Cause
Level
NIBP communication
error
NIBP measurement failure or
communication failure
Medium
Technical alarm 2
displayed in the indicating area behind NIBP MAP
:
Alarm messages
Cause
Level
SELF-TEST
FAILED
Transducer or other hardware failure.
Low
LOOSE CUFF
1. Cuff is completely unwrapped.
2. The cuff is not connected.
3. Adult cuff used in neonate mode.
AIR LEAK
Air leak in pneumatics, hose, or cuff.