Patient monitor User’s manual
51
Unable to maintain stable cuff pressure, e.g.
kinked hose.
Very weak patient signal due to a loosely
wrapped cuff. The pulse of patient is too
weak.
Measurement range exceeds module
specification.
1.Too many retries due to interference of
motion artifact.
2.Signal is too noisy during measurement,
e.g. patient has severe tremor.
3.Irregular pulse rate, e.g. arrhythmia.
Cuff pressure exceeds the specified high
safety limit. Could be due to rapid
squeezing or bumping of cuff.
Large motion artifact that saturates the BP
amplifier’s amplitude handing capability.
Module reports Air Leakage failure while in
the Pneumatic Test mode.
Module occurs abnormal processor event.
Measurement took more than 120 seconds
in adult, 90 seconds in neonate mode.
Neonate cuff used in adult mode.
Prompt messages(displayed in the indicating area behind NIBP MAP):
NIBP measurement module is resetting
NIBP is testing Over-Pressure
NIBP is testing Manometer
NIBP is testing Air Leakage
NIBP pressure calibration failed
Physiological alarm:
Sys measuring value of 〈IBP label〉
is above high alarm limit
Sys measuring value of 〈IBP label〉
is below low alarm limit
Dia measuring value of 〈IBP label〉
is above high alarm limit