Patient monitor Users manual
51
AIR PRESSURE
ERROR
Unable to maintain stable cuff pressure, e.g.
kinked hose.
WEAK SIGNAL
Very weak patient signal due to a loosely
wrapped cuff. The pulse of patient is too
weak.
RANGE
EXCEEDED
Measurement range exceeds module
specification.
EXCESSIVE
MOTION
1.Too many retries due to interference of
motion artifact.
2.Signal is too noisy during measurement,
e.g. patient has severe tremor.
3.Irregular pulse rate, e.g. arrhythmia.
OVERPRESSURE
SENSED
Cuff pressure exceeds the specified high
safety limit. Could be due to rapid
squeezing or bumping of cuff.
Medium
SIGNAL
SATURATED
Large motion artifact that saturates the BP
amplifiers amplitude handing capability.
Low
PNEUMATIC
LEAK
Module reports Air Leakage failure while in
the Pneumatic Test mode.
SYSTEM FAILURE
Module occurs abnormal processor event.
TIME OUT
Measurement took more than 120 seconds
in adult, 90 seconds in neonate mode.
CUFF TYPE ERR
Neonate cuff used in adult mode.
Prompt messagesdisplayed in the indicating area behind NIBP MAP:
Messages
Cause
Level
NIBP Resetting
NIBP measurement module is resetting
No alarm
Overpress Testing
NIBP is testing Over-Pressure
Manometer Testing
NIBP is testing Manometer
Air Leakage Testing
NIBP is testing Air Leakage
Cal. Failed
NIBP pressure calibration failed
IBP
Physiological alarm:
Alarm messages
Cause
Level
IBP labelSys high
Sys measuring value of IBP label
is above high alarm limit
Medium
User-Selectable
IBP labelSys low
Sys measuring value of IBP label
is below low alarm limit
IBP labelDia high
Dia measuring value of IBP label
is above high alarm limit