Patient monitor User’s manual
52
〈IBP label〉Dia low
Dia measuring value of 〈IBP label〉
is below low alarm limit
〈
IBP label
〉
Mean high
Mean measuring value of
〈
IBP label
〉
is above high alarm limit
〈
IBP label
〉
Mean low
Mean measuring value of
〈
IBP label
〉
is below low alarm limit
Technical alarm:
Alarm messages
Cause
Level
〈IBP label〉sensor off
〈IBP label〉sensor disconnect from
the monitor.
Low
IBP communication error
IBP communication is error
Medium
Prompt messages:
Messages
Cause
Level
〈IBP label〉zeroing
〈IBP label〉is zero calibrating
No alarm
〈
IBP label
〉
Cal. Failed
〈
IBP label
〉
pressure calibration
failed
CO
2
Physiological alarm:
Alarm messages
Cause
Level
EtCO
2
high
EtCO
2
measuring value is above high
alarm limit
Medium
User can select
Medium or
High
EtCO
2
low
EtCO
2
measuring value is below low
alarm limit
FiCO
2
high
FiCO
2
measuring value is above high
alarm limit
CO
2
Apnea
No breath detected in the set period
High
Technical alarm:
Alarm messages
Cause
Level
CO
2
sensor off
CO2 sensor off patient or off the
monitor
Low
Check airway adapter
CO2 airway adapter disconnected
with CO2 sensor
CO
2
measurement over
range
CO2 measurement Over range, need
verify zero
CO
2
sensor error
CO2 sensor error
CO
2
communication error
CO2 module failure or
communication failure
Medium