Patient monitor User’s manual
53
Sensor Over Temp
Temperature of the sensor is over the
normal working temperature.
Low
Check CO
2
Sampling Line
Sampling tube is occluded or
damaged;
Sampling tube is kinked or pinched
;
Exhaust tube is blocked.
Zero Required
The sensor needs verify zero
Sensor no initialized
Sensor or module is not initialized
Prompt messages:
Messages
Cause
Level
Zero in Progress
Zeroing is in progress.
No alarm
Sensor Warm Up
Module is warming up.
AG
Physiological alarm:
Alarm messages
Cause
Level
Et〈AG label〉high
Et〈 AG label 〉 is above high alarm
limit
Low
User-Selectable
Et〈AG label〉low
Et〈AG label〉is below low alarm limit
Fi〈AG label〉high
Fi〈 AG label 〉 is above high alarm
limit
Medium
User can select
Medium or
High
Fi〈AG label〉low
Fi〈AG label〉 is below low alarm
limit
Low
User-Selectable
EtN
2
O high
EtN
2
O is above high alarm limit
EtN
2
O low
EtN
2
O is below low alarm limit
FiN
2
O high
FiN
2
O is above high alarm limit
Medium
User-Selectable
FiN
2
O low
FiN
2
O is below low alarm limit
Low
User-Selectable
EtO
2
high
EtO
2
is above high alarm limit
Medium
User-Selectable
EtO
2
low
EtO
2
is below low alarm limit
FiO
2
high
FiO
2
is above high alarm limit
FiO
2
low
FiO
2
is below low alarm limit
Medium
User can select
Medium or
High
FiO
2
< 18%
FiO
2
measurement value is below
18%
High
Technical alarm: