Patient monitor Users manual
54
Alarm messages
Cause
Level
Check Adapter
Airway adaptor of GAS module
disconnected with sensor
Low
Unspecified accuracy
Measurement of GAS module over
range
Low
Room Air O2 Calibration
Required
Measurement of oxygen density is not
correct.
Low
Software Error
Software error.
Low
Hardware Error
Hardware error.
Low
Factory Calibration Lost
Default setting lost
Low
Motor Overspeed
Motor overspeed
Low
Replace Adapter
Please replace the adapter
Low
Zero Required
Zero calibration is required.
Low
Check Sampling Tube
Sampling tube is occluded or
damaged;
Sampling tube is kinked or pinched
Exhaust tube is blocked.
Low
O
2
Port Failure
O
2
port Failure
Low
Mix Agents
Mix Agents
Low
Prompt messages:
Messages
Cause
Level
Zeroing
Zeroing is in progress.
No alarm
ICG
Prompt messages:
Messages
Cause
Level
ICG Check Sensor
Unable to connected with modules
No Alarm
ICG Pat. Set Inval
Patient information over range
ICG Blood Set Inval
Blood pressure information over
range
ICG Ex. Measured Set
Inval
Extermal data over range
Physiological alarm: