Citadel Bed Frame System
1XUVH&DOODQG56&RQQHFWLRQV
1. Connect one end of the nurse call cable to the 37-pin D-type outlet, located
below the head end of the bed on the patient's right hand side.
2. Connect the other end of the nurse call cable to a compatible nurse call
system. The type of connector will vary depending on the nurse call system.
3. Connect one end of an RS232 cable to the 9-pin D-type outlet, located below
the head end of the bed on the patient's right hand side.
4. Connect the other end of the RS232 cable to a device capable of receiving
data through an RS232 connection.
&RQQHFWLRQWRWKHQXUVHFDOODQG56FRQQHFWRUPXVWEHPDGHZLWK
WKHFRUUHFWFDEOHV8VHRIWKHZURQJFDEOHVPD\FDXVHPDOIXQFWLRQRI
WKHEHGRUGHYLFHVFRQQHFWHGWRWKHQXUVHFDOORU56FRQQHFWRUV
Verify correct operation of the nurse call system before placing a
SDWLHQWRQWKHEHG
'RQRWDOORZWKHQXUVHFDOORU56FDEOHVWRWUDLORQWKHÀRRUZKHUH
WKH\PD\FDXVHDWULSKD]DUG8QSOXJERWKFDEOHVEHIRUHPRYLQJWKH
EHG
'RQRWDOORZWKHFDEOHVWRWRXFKWKHEHGIUDPHDVWKLVPD\DIIHFWWKH
DFFXUDF\RIWKHZHLJKLQJV\VWHP
2QO\FRQQHFWGHYLFHVGHVLJQHGWREHXVHGZLWKWKHEHG
Connecting devices not designed to be used with the bed may result
LQGDPDJHWRWKHHTXLSPHQWRUWRWKHEHG
2QO\XVHFDEOHVZLWKDQLQOLQHEUHDNDZD\FRQQHFWRUEHORZ)DLOXUH
to do so could result in damage to the bed or to other hospital
HTXLSPHQW
)LJXUH([DPSOHRIDQLQOLQHEUHDNDZD\FRQQHFWRU
Breakaway
connection
Connect to
bed
Connect to
Nurse call system