EasyManua.ls Logo

Arjo Citadel - Weighing Accuracy

Arjo Citadel
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
39
Citadel Bed Frame System
Weighing Accuracy
Patient weighing system
0LQLPXPYHUL¿FDWLRQLQWHUYDOVFDOHGLYLVLRQH 500g (1 lb)
Minimum capacity 10kg (20 lb)
Maximum capacity 270kg (595 lb)
Approvals Complies with 90/383/EEC Class III.
The weighing system is very sensitive and can be affected by a number of external
factors. For best performance, observe the following precautions:
Position the bed such that the lower frame and upper frame do not touch one
another. This may require that the bed is raised from its lowest position.
0DNHVXUHWKHEHGLVRQDÀDWOHYHOVXUIDFH
Make sure the transport locking bolts and washers have been removed before
use – refer to page 16.
Position the bed so that it is clear of obstructions. Example: walls, furniture,
cables and curtains.
Make sure the patient is fully on the mattress during weighing.
While weighing is taking place, the patient should keep as still as possible and
other persons should keep clear of the bed.
Make sure that the patient, pillows and bed linen do not touch the head board
RUÀRRUGXULQJZHLJKLQJ
The head board (1), head end lifting pole accessory sockets (2) and bed sub frame (3)
GRQRWLQÀXHQFHWKHSDWLHQWZHLJKLQJV\VWHP
$Q\LWHPDWWDFKHGWRRUSODFHGRQDQ\RWKHUSDUWRIWKHEHGZLOOLQÀXHQFHWKHSDWLHQW
weighing system. Use the AUTO COMPENSATION feature before adding or removing
any items (example: urine bags, IV poles, mattress pumps, bed-clothes, etc.).
)LJXUH$UHDQRWLQÀXHQFLQJSDWLHQWZHLJKLQJ
3