EasyManua.ls Logo

Arjo Citadel - Qualified Personnel Checks; Battery Test

Arjo Citadel
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60
Citadel Bed Frame System
$FWLRQVWREHFRPSOHWHGE\TXDOL¿HGSHUVRQQHO Yearly
Check that the bed operates correctly using the backup battery as described in the
Battery test section below
X
Check operation of the casters, paying special attention to braking and steering
functions
X
Check that the bed extension locks securely in both non-transport positions X
Examine the power cord and mains plug; if damaged, replace the complete
assembly; do not use a rewireable plug
X
([DPLQHDOODFFHVVLEOHÀH[LEOHFDEOHVIRUGDPDJHDQGGHWHULRUDWLRQ X
Check all accessible nuts, bolts and other fasteners to ensure they are present
and correctly tightened
X
&KHFNDQ\DFFHVVRULHV¿WWHGWRWKHEHGSD\LQJSDUWLFXODUDWWHQWLRQWRIDVWHQHUV
and moving parts
X
Weighing System
7KHZHLJKV\VWHPVKRXOGEHUHYHUL¿HGEHIRUHWKHGDWHRIH[SLU\VKRZQRQWKH
EHG0DNHVXUHWKHEHGLVUHYHUL¿HGHYHU\PRQWKV,IYHUL¿FDWLRQGRHVQRWSDVV
the weigh system should be re-calibrated. Refer to the product service manual for
YHUL¿FDWLRQDQGFDOLEUDWLRQSURFHGXUHVRUFRQWDFWDQ$UMRDSSURYHGVHUYLFHDJHQW
Battery Test
Check the condition of the backup battery by carrying out the following test.
1. Disconnect the bed from the electricity supply.
2. Raise the deck to maximum height - ignore the battery warning tone.
3. Raise the backrest and thigh sections to highest position.
4. 3UHVVDQGKROGWKH&35EXWWRQ7KHGHFNZLOOÀDWWHQDQGORZHUWRDPLG
height position.
5. Lower the deck to minimum height.
6. Apply maximum head down tilt (Trendelenburg).
7. Return the deck to the level position; then apply maximum foot down tilt
(reverse Trendelenburg).
If this test is not completed successfully, connect the bed to the electricity supply for at
least eight hours to recharge the battery then perform the test again. If the bed fails a
second time, contact Arjo or an Arjo-approved service agent.
To maintain best performance, the backup battery should be replaced every four years
by an Arjo-approved service agent.