34
Citadel Bed Frame System
Attendant Control Panel (ACP), Nurse Control Panel and
Patient Control Panel
Power On Indicator - Indicator LED lights when the bed is connected to
the electricity supply.
Battery Indicator - Indicates status of the battery system. Refer to the
section Backup Battery on page 32.
Anti-Entrapment System (AES) Indicator - Indicates status of the anti-
entrapment system. Refer to the section Anti-Entrapment system on page
45.
Deck Height - These buttons raise and lower the deck. All sections will
lower at maximum speed until the bed reaches low height (40cm*) where
it will continue to lower at half speed until it has reached extra low height
(32cm*). The deck will not lower to extra low height if the bed platform is
tilted (in Trendelenburg or reversed Trendelenburg position).
(*Low height and extra low height dimensions are for reference only.)
$WPLQLPXPKHLJKWFOHDUDQFHXQGHUQHDWKWKHVLGHUDLOVLVUHGXFHG
Keep your feet away from the areas below the side rails and take
H[WUDFDUHZKHQXVLQJSDWLHQWKRLVWVRUVLPLODUHTXLSPHQW
)LJXUH$WWHQGDQW&RQWURO3DQHO
)LJXUH1XUVH&RQWURO3DQHO
)LJXUH3DWLHQW&RQWURO3DQHO

Table of Contents