59
Citadel Bed Frame System
PREVENTIVE MAINTENANCE
This product is subject to wear and tear during use. To make sure that it continues to
SHUIRUPZLWKLQLWVRULJLQDOVSHFL¿FDWLRQSUHYHQWLYHPDLQWHQDQFHSURFHGXUHVVKRXOGEH
carried out at the intervals shown.
This list indicates the minimum recommended level of preventive
PDLQWHQDQFH0RUHIUHTXHQWLQVSHFWLRQVVKRXOGEHFDUULHGRXWZKHQ
the product is subjected to heavy use, aggressive environments or
ZKHUHUHTXLUHGE\ORFDOUHJXODWLRQV
Failure to carry out these checks, or continuing to use the product if
a fault is found, may compromise the safety of both the patient and
FDUHJLYHU3UHYHQWLYHPDLQWHQDQFHFDQKHOSWRSUHYHQWDFFLGHQWV
Actions to be completed by caregiver Daily Weekly
Check operation of side rails X
Visually check casters X
Check operation of the CPR backrest release handles on both sides
of the bed
X
Visually check power supply cord and mains plug X
Carry out a full test of all electrical bed positioning functions
(backrest, height, tilt, etc.)
X
Check that the patient controls, caregiver controls and Attendant
Control Panels operate correctly
X
Check that the weighing system controls operate correctly X
Check operation of the anti-entrapment system and clean the sensor
lenses (see page 45)
X
&KHFNWKHPDWWUHVVIRUGDPDJHDQGÀXLGLQJUHVV X
Examine the lifting pole, strap and handle X
If the result of any of these tests is unsatisfactory, contact Arjo or an Arjo-approved
service agent.
The procedures in the following table must be carried out by suitably
WUDLQHGDQGTXDOL¿HGSHUVRQQHO)DLOXUHWRGRVRPD\UHVXOWLQLQMXU\RU
DQXQVDIHSURGXFW