Citadel Bed Frame System
• The bed is designed to provide an acceptable side rail height when used with
a foam mattress between 15 cm (6 in) and 20.5 cm (8 in).
• Specialty powered air / foam replacement mattresses will typically envelop
the patient when loaded and can generally be deeper than a foam mattress
without compromising safety. Other makes of specialty mattress replacement
PXVWEHDVVHVVHGLQGLYLGXDOO\SULRUWRXVHWRYHULI\VXI¿FLHQWFOHDUDQFHLV
maintained.
Make sure the distance between top of side rails (if used) and top
of a specialty mattress (without compression) is approximately
FPLQRUJUHDWHU&RQVLGHULQGLYLGXDOSDWLHQWVL]H
position (relative to the top of the side rail) and patient condition
in assessing fall risk.
• To ensure compliance with EN 60601-2-52:2015, an approved Arjo mattress
should be used. Compliance with this standard when using other mattresses
must be validated by the user.
• For more information on suitable mattresses and mattress replacements,
FRQWDFW\RXUORFDO$UMRRI¿FHRUDQ$UMRDSSURYHGGLVWULEXWRU$OLVWRI$UMR
RI¿FHVFDQEHIRXQGDWWKHEDFNRIWKLVPDQXDO
6DIH6HW
• When using SafeSet with a Citadel Patient Therapy System installed, use the
extender brackets, p/n: 830.307, to extend SafeSet for improved visibility.