DISPLAY FIXED RANGE (>6% OF RANGE)
Touch Current Display (RMS)
(AC+DC)
Range 1
0.0uA ~ 32.0uA, frequency DC - 1MHz
Resolution
0.1uA
Base Accuracy
DC, 15Hz < f < 100KHz: ±(2% of
reading + 3counts)
100KHz < f < 1MHZ : ±5% of reading
(10.0uA - 999.9uA)
Range 2
7.8uA ~ 130.0uA, frequency DC -
1MHz
Resolution
0.1uA
Base Accuracy
DC, 15Hz < f < 100KHz: ±(2% of
reading + 3counts)
100KHz < f < 1MHZ : ±5% of reading
(10.0uA - 999.9uA)
Range 3
31.5uA ~ 525.0uA, frequency DC -
1MHz
Resolution
0.1uA
Base Accuracy
DC, 15Hz < f < 100KHz: ±(2% of
reading + 3counts)
100KHz < f < 1MHZ : ±5% of reading
(10.0uA - 999.9uA)
Range 4
126.0uA ~ 999.9uA, 1000uA ~
2100uA, frequency DC - 1MHz
Resolution
1uA
Base Accuracy
DC, 15Hz < f < 100KHz: ±(2% of
reading + 3counts)
100KHz < f < 1MHZ : ±5% of reading
(10.0uA - 999.9uA)
Range 5
504uA ~ 999.9uA, 1000uA ~ 8400uA,
frequency DC - 1MHz
Resolution
1uA
Base Accuracy
DC, 15Hz < f < 100KHz: ±(2% of
reading + 3counts)
100KHz < f < 1MHZ : ±5% of reading
(10.0uA - 999.9uA)
Range 6
600uA ~ 8399uA, 8.40mA ~ 10.00mA,
frequency DC 100KHz
Resolution
1uA, 0.01mA
Base Accuracy
DC, 15Hz < f < 100KHz : ±5% of
reading (0.01mA -10.00mA )

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Associated Research OMNIA II 8207 and is the answer not in the manual?

Associated Research OMNIA II 8207 Specifications

General IconGeneral
BrandAssociated Research
ModelOMNIA II 8207
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals