Chapter 4 – Functions of the instrument
UM-23100B-U User manual ATEQ F CLASS Page 70/90
4. MAINTENANCE MENU
4.1. PARAMETERS SERVICE
This menu is used to manage the memory containing the test cycle parameters.
9 Save maintenance parameters menu: used to save the configuration of the
parameters in the current test.
9 Restore maintenance parameters menu: used to restore a previously saved
configuration.
9 Erase maintenance parameters menu: used to delete the current configuration.
To access the menu, put the switch to ACCESS position.
In the main menu, place the cursor
in front of SERVICE and confirm
using ENTER.
MAIN MENU
CYCLE SPE: Disabled
CONFIGURATION
MAINTENANCE
Then place the cursor in front of
PARAMETERS and confirm using
ENTER.
MAIN/SERVICE
PARAMETERS
Then place the cursor in front of the
action required:
SAVE: save current parameters,
RESTORE: load from memory to
current parameters,
ERASE: delete current parameters
and confirm using ENTER.
IN/SERVI/PARAMETR
SAVE : No
RESTORE : No
RESET : No
To activate an operation, confirm
using ENTER. Then choose YES
using the arrows then confirm again
using ENTER.
IN/SERVI/PARAMETERS
SAVE : Yes
RESTORE : No
RESET : No
Turn the switch to LOCK position.