Chapter 3 – Start-up and settings
UM-23100B-U User manual ATEQ F CLASS Page 31/90
5.2. SETTING THE TEST PRESSURE
5.2.1. Manual setting with a mechanical regulator
Position the cursor in front of the SPE
CYCLE function and confirm with the
ENTER key.
Disabled
Regulator adjust
Infinite Fill
Next, position the cursor in front of
Regulator adjust and confirm with the
ENTER key.
Disabled
Regulator adjust
Infinite fill
The display confirms that the special
cycle has been selected.
RUN PROG. : 001
PARAMETERS
SPE CYCLE :Regul
Press the START key to launch a
special cycle.
PRESS = 355.5 mbar
REGULATOR 1 ADJUST
Set the test pressure by using the
regulator.
_
+
PRESS = 1000 mbar
REGULATOR 1 ADJUST
Once the pressure is set, press the
RESET key to stop the special cycle.
RUN/Pr:001
PRESS = 0.000 bar
READY