KBW (E6) 11/2016 page 116/147
In case of power failure, transmission of the alarm via zero-voltage relay outputs remains active for the
duration of the power failure. Afterwards, both contacts will close automatically.
When using the communication software APT-COM™ 3 DataControlSystem (option, chap.
19.1) via the Ethernet interface of the chamber for data acquisition, the alarm is not automati-
cally transmitted to the APT-COM™ protocol.
Set the tolerance limits for recording limit value excesses separately in APT-COM™ 3.
19.6 Water protected internal socket (option)
The internal socket is splash proof.
IP system of protection 67 230 V 1N ~ 50-60 Hz
Charge max. 500 W
Maximum permitted operating temperature with this option: 90 °C / 194 °F.
WARNING
Exceeding the permitted maximum temperature.
Electrical hazard.
Danger of death.
Damage to the internal socket.
∅ Do NOT exceed the temperature set-point of 90 °C / 194 °F.
Set the safety controller to 90 °C / 194 °F.
With optional temperature safety device, class 3.3, set the mechanical thermostat class
3.1 to 90 °C / 194 °F.
Heat emission of electrical devices connected inside the chamber may modify the temperature
range.
CAUTION
Risk of short circuit.
Damage to the chamber.
Use the supplied plug only (IP protection type 67). Plug it in and tighten it by screwing to
secure contact.
If the socket is not used, close the screw lid and turn it to secure.