KBW (E6) 11/2016 page 18/147
3.2 Guidelines for safe lifting and transportation
The front castors can be blocked by brakes. After operation, please observe the guidelines for temporarily
decommissioning the chamber (chap. 21.2). Please move the chambers with castors only when empty
and on an even surface, otherwise the castors may be damaged.
CAUTION
Sliding or tilting of the chamber.
Damage to the chamber.
Risk of injury by lifting heavy loads.
Transport the chamber in its original packaging only.
For moving or shipping, secure the chamber with transport straps.
∅ Do NOT lift or transport the chamber using the door, the handle, or the lower housing.
∅ Do NOT lift the chamber by hand.
Lift the chamber using technical devices (fork lifter). Set the fork lifter only from the
front or rear in the middle of the chamber.
∅ Do NOT set the fork lifter from the chamber side.
• Permissible ambient temperature range during transport: -10 °C / 14°F to +60 °C / 140°F
You can order transport packing for moving or shipping purposes from BINDER service.
3.3 Storage
Intermediate storage of the chamber is possible in a closed and dry room. Observe the guidelines for
temporary decommissioning (chap. 21.2).
• Permissible ambient temperature range during storage: -10 °C / 14°F to +60 °C / 140°F
• Permissible ambient humidity: max. 70 % r.H., non-condensing
When after storage in a cold location you transfer the chamber to its warmer installation site, condensa-
tion may form. Before start-up, wait at least one hour until the chamber has attained ambient temperature
and is completely dry.
3.4 Location of installation and ambient conditions
Set up the chamber on a flat, even surface, and in a well-ventilated, dry location and align it using a spirit
level. The site of installation must be capable of supporting the chamber’s weight (see technical data,
chap. 23.4). The chambers are designed for setting up inside a building (indoor use).
Danger of overheating.
Damage to the chamber.
∅ Do NOT set up chamber in non-ventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.