KBW (E6) 11/2016 page 72/147
Submenu “Various”.
Select the field ”Temperature range”.
-
“Temperature range” entry menu.
Enter the desired value for the temperature
range and press the Confirm icon.
Setting range: 2 °C to 20 °C
There are the following possibilities:
Press the Confirm icon to take over the entries and exit the menu.
Press the Close icon to exit the menu without taking over the entries.
If there are actual values outside the tolerance range the following information icons for the corresponding
parameter are displayed:
Temperature range
The temperature value is outside the tolerance range
If the condition persists, an alarm is triggered after the configured interval (“range alarm delay”). It is visu-
ally indicated in Normal display. If the alarm buzzer is activated (chap. 11.5) there is an audible alert. The
zero-voltage relay alarm output (option, chap. 19.5) switches to transmit the alarm. The alarm is shown in
the list of active alarms (chap. 11.3).
11.5 Activating / deactivating the audible alarm (alarm buzzer)
Path: Main menu > Settings > Chamber
“Chamber” submenu.
In the field “Audible alarm” select the desired setting
“off“ or “on” and press the Confirm icon.