KBW (E6) 11/2016 page 128/147
Fault description
Possible cause
Required measures
Refrigerating performance
Low or no refrigerating perfor-
mance.
Ambient temperature > 25 °C / 77
°F (chap.3.4).
Select cooler place of installation.
Compressor not turned on.
Contact BINDER service.
Electro-valves defective.
No or not enough refrigerant.
Too much external heat load.
Reduce heat load.
Icing at the evaporator plates.
Set-point was too long below
ambient temperature.
Defrost the chamber (chap. 17).
Condensation at the walls of the
inner chamber.
Set-point was too long below
ambient temperature, icing in the
preheating chamber.
Defrost the chamber (chap. 17)
Controller
No chamber function
(dark display).
Display in standby mode.
Press on touchscreen.
Main power switch turned off.
Turn on the main power switch.
Menu functions not available.
Menu functions not available with
current authorization level.
Log in with the required higher
authorization.
No access to controller
Password incorrect.
Contact BINDER service.
Chart recorder function: meas-
ured-value memory cleared; in-
formation lost.
New setting of storage rate or
scaling (minimum and/or maxi-
mum) (chap. 16.2).
Change the storage rate or scal-
ing ONLY if the previously regis-
tered data are no longer required.
Controller does not equilibrate to
setpoints entered in Fixed value
operation mode
Controller is not in Fixed value
operation mode.
Change to Fixed value operation
mode.
Controller does not equilibrate to
program set-points.
Controller is not in program oper-
ation mode, or program delay
time is running.
Start the program again. If appro-
priate, wait for the program delay
time.
Program duration longer than
programmed.
Tolerances have been pro-
grammed.
For rapid transition phases, do
NOT program tolerance limits in
order to permit maximum heating
or refrigerating speed.
Program keeps the last program
setpoint constant while in setting
“ramp”.
Program line with setting “ramp”
is incomplete.
When programming with setting
“ramp”, define the end value of
the desired cycle by adding an
additional section with a section
time of at least one second.
Ramp temperature transitions are
only realized as steps.
Setting “step” has been selected. Select setting “ramp”.
Acknowledging the alarm does
not cancel the alarm state.
Cause of alarm persists.
Remove cause of alarm. If the
alarm state continues, contact
BINDER service.
Alarm message:
- - - - or <-<-< or >->->
Sensor rupture between sensor
and controller or Pt 100 sensor
defective.
Contact BINDER service.
Short-circuit.
Miscellaneous
Fluorescent tube does not illumi-
nate.
Defective fluorescent tube. Replace the fluorescent tubes.
Only qualified service personnel authorized by BINDER must perform repair.
Repaired chambers must comply with the BINDER quality standards.