EasyManua.ls Logo

Binder CB Series - CB; CB-UL: Humidity Control of the Permadry System

Binder CB Series
199 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CB / CB-UL, CBF / CBF-UL (E7) 01/2022 Page 72/199
When operating the CBF / CBF-UL with deactivated humidity control, humidification can be realized using
the Permadry water pan (chap. 6.8).
6.8 CB / CB-UL: Humidity control of the Permadry™ system
With the CBF / CBF-UL, humidity setting via the Permadry System is not required.
The Permadry™ system with its 2-pan water system guarantees a maximum humidity of up to 95 % r.H. in
the inner chamber, which remains condensation-free. This performance assumes an average ambient
temperature of 22 +/- 3 °C / 71.6 +/- 5.4 °F and a working temperature in the inner chamber of 37 °C /
98.6 °F.
You can increase or decrease humidity slightly via the heating of the water pan if required.
Humidity is infinitely adjustable in whole numbers from -5 (low) to +5 (high).
Factory setting: “0” (medium).
Setting:
Path: Main menu > Settings > Various
Submenu “Various”.
Select the field “Humidity control”
“Humidity control” entry menu. Enter the desired
humidity and confirm entry with Confirm icon.
If the ambient temperature deviates by more than +/- 5 °C from the values recommended by
the manufacturer, the conditions for maximum air humidity with condensation-free inner
chamber are no longer guaranteed. Contact BINDER Service for assistance.

Table of Contents