102
10. Select the Continue icon. The attachment will move automatically to the Horizontal
position.
11. Watch the video. Attach the calibration weight horizontally and allow it to settle.
12. Select the Continue icon.
13. Verification will occur if torque is within the acceptable tolerance range.
14. Click the Report button to get the Verification History report. The report for previous
verification can be accessed anytime by clicking the Report button.
NOTE: A Verification Failure message will display if torque is not within the acceptable
tolerance range. If the system will not verify within an acceptable range, please contact
Biodex Service (see contact information).
NOTE: For research applications, the user may want to perform the Calibration
Verification after a specific number of patients in case questions regarding the validity
of the data arise at a later date.
NOTE: The attachment will not be released and will not proceed to the next step when
weight is detected.
NOTE: Additional information can be found at: www.biodex.com/s4
Disposal
An appropriate waste disposal company is to be contacted (i.e., the local collection point for
waste separation). Properly dispose of the device at the end of its service life:
The device packaging is disposed of through resource recycling.
The metal parts of the machine go to scrap metal disposal.
Plastic parts are disposed of as hazardous waste.
NOTE: The disposal of equipment must be in accordance with the respective
national regulations.
Wear parts are considered hazardous waste! After being replaced, wear
parts must be disposed of according to country-specific waste laws.