Handheld monitor user’s manual
15
Guidance and manufacture’s declaration electromagnetic immunity for all
EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration electromagnetic immunity
The M800 HANDHELD MONITOR is intended for use in the electromagnetic environment
specified below. The customer or the user of M800 HANDHELD MONITOR should assure
that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood,
concrete or ceramic tile.
If floor are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for signal
line
0.5 kV for power
supply lines
0.5 kV for signal
line
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
1 kV
differential mode
2 kV common
mode
1 kV differential
mode
2 kV common
mode
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
Mains power quality
should be that of a
typical commercial or
hospital environment.