Handheld monitor user’s manual
17
Immunity
test
IEC 60601 test
level
Complian
ce level
Electromagnetic environment -
guidance
Conducted
RF
IEC
61000-4-6
Radiated RF
IEC
61000-4-3
3 V
rms
150 kHz to 80
MHz
3 V/m
80 MHz to 2.5
GHz
1 V
rms
3 V/m
Portable and mobile RF
communications equipment should be
used no closer to any part of the
M800 HANDHELD MONITOR,
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
P
V
d
1
5.3
P
E
d
1
5.3
80 MHz to 800 MHz
P
E
d
1
7
800 MHz to 2.5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the
recommended separation distance in
metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should
be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity
of equipment marked with the
following symbol: