N-19Rev F
InStallatIon
N-19
patIent reSult approval
Some institutions may desire patient test results to be verified by the user before printing
or sending to the LIS. Activation of this feature displays the result, then requires the user
to accept or reject the result before the record is printed and saved. If the test is rejected,
the meter will permanently flag it as rejected by the user.
1. From a list of programmable parameters, choose meter SettIngS
using the
keys.
2. Press the key.
3. Use the
keys to select
comm SettIngS.
4. Press the key.
5. Use the
keys to select
pat. reSult approval.
6.
Use the
keys to
choose dISaBled or enaBled.
7. Press the key.
When enabled, after a patient result is available and displayed on the meter screen, the
user must accept or reject the result before printing, uploading or performing any other
meter functions.
The meter prompts the user to preSS 1 to accept or 0 to reject. Once the test has been
accepted or rejected, the meter operates normally.
If the test has been rejected, the display and printout will be flagged with pat. reSult
rejected or reSultS rejected By uSer. Recalled results will display an r before
the patient ID in the rejected record.
maIn menu
Set parameterS
meter SettIngS
comm SettIngS