28 | Chapter 4: Instrument Qualification
Absorbance Test Plate Results
Reader: Epoch (Serial Number: 182843)
Basecode: P/N 7200201 (v1.07)
Date and Time: 7/14/2009 8:49:39 AM
Absorbance Plate: 7 Filter Test Plate (P/N 7260522) - S/N 161259
Last Plate Certification: April 2009
Next Plate Certification Due: April 2010
User: Administrator
Comments:
Peak Absorbance Results
Well C6
Reference 586
Tolerance 3
Read 586
Result PASS
Alignment Results
Wells B2 B12 G1 G11
Read 0.000 0.000 0.000 0.000
Tolerance 0.015 0.015 0.015 0.015
Result PASS PASS PASS PASS
Wavelength = 405 nm
Accuracy Results
Wells C1 E2 G3 H6 F5 D4
Reference 0.147 0.618 1.133 1.701 2.279 2.945
Min Limit 0.124 0.586 1.090 1.647 2.168 #N/A
Max Limit 0.170 0.650 1.176 1.755 2.390 #N/A
Read 1 0.143 0.614 1.129 1.693 2.273 2.902
Result PASS PASS PASS PASS PASS #N/A
Repeatability Results
Wells C1 E2 G3 H6 F5 D4
Read 1 0.143 0.614 1.129 1.693 2.273 2.902
Min Limit 0.137 0.603 1.113 1.671 2.200 #N/A
Max Limit 0.150 0.625 1.145 1.715 2.346 #N/A
Read 2 0.143 0.614 1.129 1.694 2.269 2.918
Result PASS PASS PASS PASS PASS #N/A
Wavelength = 450 nm
Accuracy Results
Wells C1 E2 G3 H6 F5 D4
Reference 0.140 0.575 1.052 1.578 2.024 2.604
Min Limit 0.117 0.543 1.011 1.526 1.923 #N/A
Max Limit 0.163 0.606 1.093 1.630 2.125 #N/A
Read 1 0.136 0.571 1.047 1.570 2.016 2.591
Result PASS PASS PASS PASS PASS #N/A
Repeatability Results
Wells C1 E2 G3 H6 F5 D4
Read 1 0.136 0.571 1.047 1.570 2.016 2.591
Min Limit 0.130 0.560 1.031 1.549 1.951 #N/A
Max Limit 0.143 0.581 1.062 1.590 2.081 #N/A
Read 2 0.137 0.571 1.047 1.570 2.017 2.589
Result PASS PASS PASS PASS PASS #N/A
BioTek
®
Instruments, Inc.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Biotek Epoch and is the answer not in the manual?

Biotek Epoch Specifications

General IconGeneral
TypeMicroplate Spectrophotometer
Wavelength Range200 - 999 nm
Light SourceXenon flash lamp
Power Requirements100 - 240 VAC, 50/60 Hz
Wavelength Repeatability+/- 0.2 nm
Absorbance Range0-4.0 OD
Plate Types6- to 384-well microplates
ShakingLinear, orbital
Detection ModesAbsorbance
SoftwareGen5
Bandwidth5 nm
ConnectivityUSB