Absorbance Plate Test | 29
Wavelength = 490 nm
Accuracy Results
Wells C1 E2 G3 H6 F5 D4
Reference 0.135 0.574 1.051 1.577 1.976 2.545
Min Limit 0.112 0.543 1.010 1.525 1.916 #N/A
Max Limit 0.158 0.605 1.092 1.629 2.036 #N/A
Read 1 0.132 0.571 1.048 1.571 1.973 2.537
Result PASS PASS PASS PASS PASS #N/A
Repeatability Results
Wells C1 E2 G3 H6 F5 D4
Read 1 0.132 0.571 1.048 1.571 1.973 2.537
Min Limit 0.125 0.560 1.033 1.551 1.948 #N/A
Max Limit 0.138 0.582 1.063 1.592 1.998 #N/A
Read 2 0.131 0.571 1.048 1.571 1.974 2.540
Result PASS PASS PASS PASS PASS #N/A
Wavelength = 550 nm
Accuracy Results
Wells C1 E2 G3 H6 F5 D4
Reference 0.130 0.568 1.040 1.560 1.956 2.513
Min Limit 0.107 0.537 0.999 1.509 1.897 #N/A
Max Limit 0.153 0.599 1.081 1.611 2.015 #N/A
Read 1 0.127 0.566 1.038 1.556 1.955 2.510
Result PASS PASS PASS PASS PASS #N/A
Repeatability Results
Wells C1 E2 G3 H6 F5 D4
Read 1 0.127 0.566 1.038 1.556 1.955 2.510
Min Limit 0.121 0.555 1.023 1.535 1.931 #N/A
Max Limit 0.134 0.576 1.053 1.577 1.980 #N/A
Read 2 0.127 0.565 1.038 1.556 1.954 2.505
Result PASS PASS PASS PASS PASS #N/A
Wavelength = 620 nm
Accuracy Results
Wells C1 E2 G3 H6 F5 D4
Reference 0.136 0.573 1.050 1.575 1.893 2.437
Min Limit 0.113 0.542 1.009 1.523 1.835 2.320
Max Limit 0.159 0.604 1.091 1.627 1.951 2.554
Read 1 0.133 0.571 1.049 1.572 1.894 2.430
Result PASS PASS PASS PASS PASS PASS
Repeatability Results
Wells C1 E2 G3 H6 F5 D4
Read 1 0.133 0.571 1.049 1.572 1.894 2.430
Min Limit 0.127 0.560 1.033 1.551 1.870 2.352
Max Limit 0.140 0.582 1.064 1.592 1.918 2.508
Read 2 0.133 0.571 1.049 1.571 1.895 2.430
Result PASS PASS PASS PASS PASS PASS
Wavelength = 630 nm
Accuracy Results
Wells C1 E2 G3 H6 F5 D4
Reference 0.136 0.568 1.040 1.560 1.865 2.400
Min Limit 0.113 0.537 0.999 1.509 1.808 2.284
Max Limit 0.159 0.599 1.081 1.611 1.922 2.516
Read 1 0.134 0.566 1.039 1.557 1.864 2.392
Result PASS PASS PASS PASS PASS PASS
Epoch Operator’s Manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Biotek Epoch and is the answer not in the manual?

Biotek Epoch Specifications

General IconGeneral
TypeMicroplate Spectrophotometer
Wavelength Range200 - 999 nm
Light SourceXenon flash lamp
Power Requirements100 - 240 VAC, 50/60 Hz
Wavelength Repeatability+/- 0.2 nm
Absorbance Range0-4.0 OD
Plate Types6- to 384-well microplates
ShakingLinear, orbital
Detection ModesAbsorbance
SoftwareGen5
Bandwidth5 nm
ConnectivityUSB

Related product manuals