28 en | Programming NurseCall Main Unit
F.01U.252.699 | V1.1 | 2012.01 User Manual Bosch Security Systems
Example of programming
Hereafter is an example of the programming, with the following characteristics:
Locating mode = ON
Display mode = FL/RO/BE
Protocol = ESPA 4.4.4
RPE 670 / i-page system = YES
Day / night transfer function = YES
See Section A.5 Paging systems specifications, page 53.
1. Selected PAGING in the RS-232 Output menu. Confirm with .
2. Select the protocol (ESPA 4.4.4, POCSAG, DeTeWe or Medicall 800). Confirm with .
3. Activate (YES) or deactivate (NO) the RPE 670 system with and .
4. Confirm with .
5. Enter the ID number of the paging system (0-9) and the NurseCall system (0-9).
Select a field with and , change a value with and . Confirm with .
6. Activate (YES) or deactivate (NO) the day / night transfer with and .
7. Confirm with .
8. If the day / night transfer is activated, set the night start time. Default value is 18:00.
Select a field with and , change a value with and . Confirm with .
9. Set the night end time and confirm your setting with . Default value is 6:00.
¬ RS-232 Output
PAGING ¡
¬ PAGING
ESPA 4.4.4. ¡
¬ PAGING
RPE 670 ? YES ¡
¬ PAGING ESPA 4.4.4.
ID PAG.:2 ID NCALL:1
¬ PAGING ESPA 4.4.4.
Pager DAY-NIGHT YES¡
¬ PAGING ESPA 4.4.4.
Night Beg.: 18:00:00
¬ PAGING ESPA 4.4.4.
Night End : 06:00:00