Brookeld Engineering Labs., Inc. Page 34 Manual No. M/03-165-D0410
2. For the Time To Torque method, the DV-II+Pro viscometers will retain the last
entered torque value for use when next the user elects to perform a time to torque test.
3. The user can set up a desired print interval time, then set the viscometer to the
continuous print mode and nally initiate either of the timed modes of operation.
While setting up the timed mode parameters, continuous print operation will cease.
However, upon starting the timed operation, the DV-II+Pro will output an initial data
string to the printer and then continue printing data strings (at the user dened time
interval) for the duration of the timed run. At the end of the timed run, continuous
printing will again be disabled and the user may print single strings (of the nal data
point) at your option until you exit the timed mode. Upon returning to the default
operation mode, continuous printing will again resume at the user selected time
interval. In a similar manner, if you are in the once-per-PRINT -key-press mode, when
you enter the timed mode of operation you will be able to print data strings at any
time during the timed mode by pressing the PRINT key.
III.3.3 Print Time Interval
This option is used to set the print time interval to the selected printer. Scroll to Set Print Time,
as shown in Figure III-28:
D TIMED TO TORQUE è
SETPRINTTIME ê
Figure III-28
Press the ENTER/AUTORANGE key. On entry, Figure III-29 is displayed:
PRINT INTERVAL
SET MIN’S: 00 èê
Figure III-29
Using the UP and DOWN ARROW keys, enter a value for the minutes between successive print
strings. This value can be as high as fty-nine (59) minutes and as low as 00.
When satised, press the OPTIONS/TAB key to enter the seconds setting display:
PRINT INTERVAL
SET SEC’S: 00 èê
Figure III-30
Using the UP and DOWN ARROW keys, enter a value for the seconds portion of the print interval.
This value can be between zero (0) and fty-nine (59 )seconds.