EasyManua.ls Logo

Brookfield RVDV-II+Pro - Appendix J - DVE-50 A Probe Clip

Brookfield RVDV-II+Pro
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Brookeld Engineering Labs., Inc. Page 81 Manual No. M/03-165-D0410
Appendix J - DVE-50A Probe Clip
Probe Clip DVE-50A is supplied with all model DV-II+Pro Viscometers, DV-III
Rheometers, and Digital Temperature Indicators. It is used to attach the RTD temperature
probe to the LV Guard Leg (Part No. B-20Y) or 600 ml low form Grifn beaker. Figure J1
is a view of the Probe Clip, showing the hole into which the RTD probe is inserted, and the
slot which ts onto the LV guard leg. When inserting the RTD probe into the Probe Clip,
the upper part of the Clip is compressed by squeezing the points shown in Figure J1.
Note: All Viscometer/Rheometer models except LV use the Probe Clip as
shown in Figure J3.
RTD Probe
Hole
Squeeze Here when
installing RTD
Temperature probe
Figure J1
Figure J2 shows the Probe Clip (with RTD temperature probe installed) mounted on the LV
guard leg.
Figure J3 shows the Probe Clip mounted in a 600 ml low form Grifn beaker. This
mounting may be used with LV, RV, HA and HB series instruments.
Note: The RTD probe must be parallel to the beaker wall so as not to interfere with the
viscosity measurement.
Figure J2
Figure J3
6
0
0 ml
500
ml
400
300
200
100
USA
No. 14000
600 ml Low Form
Griffin Beaker
DVE–50
Probe Clip
RTD Temperature Probe
LV Guard Leg
Assembly B-20KY
Probe Clip
DVE-50
R TD Temperatur e
Probe

Table of Contents

Related product manuals