Brookeld Engineering Labs., Inc. Page 35 Manual No. M/03-165-D0410
Note: The value for minutes or seconds must be other than zero (0) or you will print
continuously when you exit this mode. A press of the MOTOR ON/OFF/
ESCAPE key would exit this option and take you back to the screen of Figure
III-28.
Continued pressing of the OPTIONS/TAB key will toggle between the minutes and seconds input
screens. Press the ENTER/AUTORANGE key to accept the new values for print interval in minutes
and seconds. You will now be in the screen display of Figure III-28 where you may re-enter
the print interval mode, or exit to the default screen (Figure II-5) by pressing the MOTOR ON/
OFF/ESCAPE key.
Activating print selections in the Print mode can only be done by exiting to the main menu
and pressing the PRINT key for four (4) seconds. “P%” will ash in front of the torque reading,
conrming that you are now in the Print Interval mode. Pressing PRINT for one (1) second
thereafter will disable the Print mode and remove the “P%” from the display.
III.3.4 PC Program (On/O)
This option causes the serial port of the DV-II+Pro viscometer to go into a high speed output
mode (approximately 3 print lines per second) for use with Brookeld WINGATHER®
software program. When ON, you may enter the Options menu but will not be allowed to
make any option selections until the PC PROG is turned OFF. All front panel keys will
function normally when you turn the option ON and return to normal viscometer operation by
pressing the MOTOR ON/OFF/ESCAPE key. When OFF, the DV-II+Pro will return to the last set
print time interval when printing is resumed.
Note:
WINGATHER® software can be used for data acquisition and analysis.
Rheocalc32 software can be used for complete external control of DV-
II+Pro.
From the Options menu, scroll to the screen shown in Figure III-31:
SET PRINT TIME è
D PCPROGOFF ê
Figure III-31
Press the OPTIONS/TAB key to display Figure III-32:
SET PRINT TIME è
D PCPROGON
ê
Figure III-32
Pressing the OPTIONS/TAB key would return you to the screen display of Figure III-31. Repeated
pressing of the OPTIONS/TAB key would cause you to toggle back-and-forth between the displays
of Figure III-31 and Figure III-32.
To turn high speed output ON, press the ENTER/AUTORANGE key when the appropriate screen is
displayed. Then press the MOTOR ON/OFF/ESCAPE key to exit the Setup mode. This returns you