Brookeld Engineering Labs., Inc. Page 38 Manual No. M/03-165-D0410
USE SPINDLE 31
PRESS A KEY
Figure III-40
where the operator is instructed to mount a spindle 31 and then press a key to continue. Or you
might see:
SET TEMP TO 100C
PRESS A KEY
Figure III-41
followed by:
WAIT TEMP 100C
P1 S00/12 00:14
Figure III-42
Here, the B.E.V.I.S. program is waiting for the temperature to reach one hundred (100) °C
before it continues to the next program step. Also displayed is the program number (P1), the
step number and the total program steps (S00/12) and the elapsed time since the program (or
step) began (00:14). While in this “wait state” you can press the OPTIONS/TAB key to see the
viscometer default screen in Figure III-43; the PGM will be ashing.
CP 123.4 55.1C
10 RPM PGM% 63.7
Figure III-43
The DVLOADER Software is used to create, save, print, and down load programs to the DV-
II+Pro Viscometer. The next section explains how to use the DVLOADER software.