Brookeld Engineering Labs., Inc. Page 59 Manual No. M/03-165-D0410
Special Considerations
In taking viscosity measurements with the DV-II+Pro Viscometer there are two considerations
which pertain to the low viscosity limit of effective measurement.
1) Viscosity measurements should be accepted within the equivalent % Torque Range
from 10% to 100% for any combination of spindle/speed rotation.
2) Viscosity measurements should be taken under laminar ow conditions, not under
turbulent ow conditions.
The rst consideration has to do with the precision of the instrument. All DV-II+Pro
Viscometers have an accuracy of +/- 1% of the range in use for any standard spindle or cone/
plate spindle. (Note that accuracy values may be higher than 1% when using accessory devices
with the DV-II+Pro). We discourage taking readings below 10% of range because the potential
viscosity error of +/- 1% is a relatively high number compared to the instrument reading.
The second consideration involves the mechanics of uid ow. All rheological measurements
of uid ow properties should be made under laminar ow conditions. Laminar ow is ow
wherein all particle movement is in layers directed by the shearing force. For rotational
systems, this means all uid movement must be circumferential. When the inertial forces on
the uid become too great, the uid can break into turbulent ow wherein the movement of
uid particles becomes random and the ow can not be analyzed with standard math models.
This turbulence creates a falsely high Viscometer reading with the degree of non-linear
increase in reading being directly related to the degree of turbulence in the uid.
For the following geometries, we have found that an approximate transition point to turbulent
ow occurs:
1) No. 1 LV Spindle: 15 cP at 60 RPM
2) No. 1 RV Spindle: 100 cP at 50 RPM
3) UL Adapter: 0.85 cP at 60 RPM
Turbulent conditions will exist in these situations whenever the RPM/cP ratio exceeds the values
listed above.
Eect on accuracy when using accessory devices
The Brookeld viscometer has a stated accuracy of +/- 1% of the range in use. This stated
accuracy applies when the viscometer is used in accordance with the operating instructions
detailed in the instrument instruction manual and the calibration test uid is used in accordance
with the instructions provided by the uid supplier (including the critical parameters of
temperature control and stated uid accuracy). Brookeld’s accuracy statement of +/- 1% of
the range in use applies to the Brookeld rotational viscometer when used with the standard
spindles supplied with the instrument, including LV spindles 1 through 4 (supplied with LV
series viscometers), RV spindles 2 through 7 (supplied with RV series viscometers), and HV
series spindles 2 through 7 (supplied with HA series viscometers and HB series viscometers) in
a 600 ML low form Grifn beaker.
Brookeld offers a range of accessories for use with the Brookeld viscometer to
accommodate special measurement circumstances. These accessories, while offering added
capability to the user, also contribute to an expanded measurement tolerance beyond the
instrument accuracy of +/- 1% of the range in use. This expanded measurement tolerance is
a function of many parameters including spindle geometry, accessory alignment accuracy,
sample volume requirement, and sample introduction techniques. The effect of these