Brookeld Engineering Labs., Inc. Page 67 Manual No. M/03-165-D0410
Brookeld Viscosity Standard Fluid General Information
We recommend that Brookeld Viscosity Standard Fluids be replaced on an annual basis, one
year from date of initial use. These uids are pure silicone and are not subject to change over
time. However, exposure to outside contaminants through normal use requires replacement on
an annual basis. Contamination may occur by the introduction of solvent, standard of different
viscosity or other foreign material.
Viscosity Standard Fluids may be stored under normal laboratory conditions. Disposal should
be in accordance with state, local and federal regulations as specied on the material safety
data sheet.
Brookeld Engineering Laboratories does not recertify Viscosity Standard Fluids. We will
issue duplicate copies of the Certicate of Calibration for any uid within two years of the
purchase date. Brookeld Viscosity Standard Fluids are reusable provided they are not
contaminated. Normal practice for usage in a 600 ml beaker is to return the material from the
beaker back into the bottle. When using smaller volumes in accessories such as Small Sample
Adapter, UL Adapter or Thermosel, the uid is normally discarded.
Calibration Procedure for LV(#1-4) and RV,HA,HB(#1-7) Brookeld Spindles
1) Place the viscosity standard uid (in the proper container) into the water bath.
2) Lower the DV-II+Pro into measurement position (with guard leg if LV or RV series Viscometer
is used).
3) Attach the spindle to the Viscometer. If you are using a disk-shaped spindle, avoid trapping
air bubbles beneath the disk by rst immersing the spindle at an angle, and then connecting it
to the Viscometer.
4) The viscosity standard uid, together with the spindle, should be immersed in the bath for a
minimum of 1 hour, stirring the uid periodically, prior to taking measurements.
5) After 1 hour, check the temperature of the viscosity standard uid with an accurate
thermometer.
6) If the uid is at test temperature (±0.1°
C of the specied temperature, normally 25°C), measure
the viscosity and record the Viscometer reading. Note: The spindle must rotate at least ve (5)
times before readings are taken.
7) The viscosity reading should equal the cP value on the uid standard to within the combined
accuracies of the Viscometer and the viscosity standard (as discussed in the section entitled,
Interpretation of Calibration Test Results) which appears later in this section.
Calibration Procedure for a Small Sample Adapter
Brookeld recommends a two step check. First verify the calibration of the viscometer using
the standard viscometer spindles (LV1-4, RV2-7, HA2-7 and HB2-7 or cone/plate spindles) as
detailed in this appendix. Second verify the calibration of the viscometer using the Small Sample
Adapter. The use of an accessory device may increase the accuracy of measurement associated
with the DV-II+Pro.