Brookeld Engineering Labs., Inc. Page 84 Manual No. M/03-165-D0410
No Recorder Response
Be Sure the viscometer is not at ZERO reading.
Be sure the recorder is ON and not on STANDBY.
Verify the range settings.
Check cable leads for clean connection.
Recorder Pen Moves in Wrong Direction
Output polarity reversed
Reverse leads
Viscometer Will Not Communicate with PC
Check the comm port and make sure the correct port is being utilized.
Check the interconnecting cable for proper installation
Check the Options menu and make sure the PC PROG is set to either “ON” or “OFF” in
accordance with the operating instructions for the program/procedure in use.
If the above do not rectify the problem, do the following:
Shut off viscometer
Attach printer.
Press and hold the MOTOR ON/OFF and ENTER/AUTORANGE keys simulanteously while
turning the viscometer power on. Figure K1 appears on the viscometer display.
VISCOMETER SETUP
ENTER TO START
Figure K1
Press ENTER/AUTORANGE key and Figure J2 appears.
PRINT PARAM RAM?
NO#$ THEN ENTER
PRINT PARAM RAM?
YES#$ THEN ENTER
"
t
Figure K2