Magnetic Stray Fields
40 H157655_1_008
Please note all measurements in the following tables are in meters!
6.1 Horizontal Stray Fields
Magnet Type 5.0 mT
(50 G)
3.0 mT
(30 G)
1.0 mT
(10 G)
0.5 mT
(5 G)
0.2 mT
(2 G)
0.1 mT
(1 G)
0.05 mT
(0.5 G)
400/54 Ascend Aeon 0.38 0.41 0.46 0.50 0.65 0.83 1.08
500/54 Ascend Aeon 0.47 0.50 0.56 0.60 0.74 0.96 1.26
600/54 Ascend Aeon 0.52 0.56 0.62 0.70 0.92 1.18 1.55
700/54 Ascend Aeon 0.49 0.52 0.65 0.80 1.10 1.42 1.87
Distances are measured in radial direction from magnetic center.
Table6.2: Horizontal Stray Fields: Standard Bore 54 mm
Magnet Type 5.0 mT
(50 G)
3.0 mT
(30 G)
1.0 mT
(10 G)
0.5 mT
(5 G)
0.2 mT
(2 G)
0.1 mT
(1 G)
0.05 mT
(0.5 G)
400/89 Ascend Aeon 0.44 0.47. 0.55 0.60 0.69 0.82 1.03
500/89 Ascend Aeon 0.50 0.54 0.62 0.70 0.92 1.17 1.52
600/89 Ascend Aeon 0.60 0.64 0.76 0.85 1.07 1.33 1.71
700/89 Ascend Aeon 0.80 0.88 1.09 1.25 1.50 1.81 2.26
Distances are measured in radial direction from magnetic center.
Table6.3: Horizontal Stray Fields: Wide Bore 89 mm

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bruker AVANCE and is the answer not in the manual?

Bruker AVANCE Specifications

General IconGeneral
BrandBruker
ModelAVANCE
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals