CryoProbe and Other Accessories
H157655_1_008 95
10.6 CryoFit
This section describes the requirements for the site planning for the CryoFit.
10.6.1 Introduction
A CryoFit converts a standard CryoProbe into a flow probe within a few minutes. The CryoFit
is  a  long  metal  pole  with  a  flow  cell.  The  flow  cell  is  positioned  inside  the  CryoFit  during
storage and  installation. The  outlet  capillary is  attached to  the flow  cell  and runs  inside the
metal-pole to the top of the CryoFit. The inlet capillary is attached to the flow cell and is guide
during the installation procedure through the VT-air inlet of the probe to the outside.
CryoFits  are  compatible  with  all  newer  5  mm  He-cooled  CryoProbes  as  well  as  Prodigy
CryoProbes.
The  CryoFit  is  inserted  from  the  top  of  the  magnet  into  the  upper  part  of  the  shim  system
(BST Upper  Part)  is pushed  into  the magnet  until  the end  of  the CryoFit  sits  in the  spinner
turbine. After this installation procedure the flow cell is pushed into the NMR probe so that is
located in the area of the NMR coils like a sample tube.
10.6.2 Installation Requirements
Two essential requirements have to be observed for the CryoFit.
1. The CryoFit must be long enough to that it fits into the BST and the flow cell can reach
the magnet center. This requirement is automatically checked during the order review.
– If a CryoFit is ordered as accessory for an already existing, older magnet the length
should be verified as the automatic conversion might not work properly.
– The automatically selected CryoFits are will be compatible with all types of sample
changers and included some security margin. In case a problem with the available
ceiling height –see (2) below–, you can verify if the smaller CryoFit fits the current
instrument configuration, for example if no SampleChanger is included.
2. The CryoFit requires a minimum ceiling height so that the distance between BST upper
part and the ceiling allows the insertion of the CryoFit.
– The calculation of the required ceiling height depends on many parameters (type of
magnet, magnet feet, BST length, type of sample changer).
For  the  calculation  of  the  magnet  ceiling  height  (MCH)  refer  to  the  CryoFit  Site  Planning
Guide.