CryoProbe and Other Accessories
H157655_1_008 83
10 CryoProbe and Other
Accessories
The CryoProbe accessory for the AVANCE
Series NMR Spectrometers offers dramatic
increases in signal to noise ratio (S/N) by reducing the operating temperature of the NMR coil
assembly and the preamplifier.
The CryoProbe accessory consists of two major components: the CryoProbe and the
CryoPlatform. The CryoProbe is similar to a standard probe, however contains cryogenically
cooled RF coil and electronics. The CryoPlatform, which provides cryogenic cooling for the
CryoProbe, is made up of:
CryoCooling Unit with control electronics.
Helium Compressor (along with any associated cooling equipment).
Helium Gas Cylinder (for purging of the CryoProbe).
Helium transfer lines and transfer line support.
The CryoCooling Unit must be sited adjacent to the magnet, whereas the Helium Compressor
and Helium Gas Cylinder may be sited remotely. The first step in the site planning involves
review of the magnet area to determine if floor space is available to accommodate the
CryoCooling Unit. The CryoCooling Unit is optimally located level with the magnet. Other
configurations, such as magnet pits that can not accommodate the CryoCooling Unit, require
special consideration.