CryoProbe and Other Accessories
88 H157655_1_008
10.2.1.2 Helium Compressor - Indoor Air Cooled
The room air handling system must be able to dissipate 7.5 kW of heat.
Siting this helium compressor in the same room as the NMR is not recommended.
Ambient operating temperature 5–28ºC.
Figure10.5: Helium Compressor - Indoor Air Cooled
10.2.1.3 Helium Compressor - Outdoor Air Cooled
The Outdoor Air Cooled helium compressor consists of an Outdoor Unit and an Indoor
Unit.
The outdoor unit is specified to operate between the temperatures of –30ºC to 45ºC.
Figure10.6: Helium Compressor - Outdoor Air Cooled