It is recommended that you use Super Sani-Cloth
®
Germicidal Disposable Wipes by PDI, Inc. or bleach (0.6%
Sodium Hypochlorite) and clean with non-linting wipes.
To disinfect the probe using the Intermediate-Level Disinfection (ILD) method with Super Sani-Cloth
®
Germicidal Disposable Wipes by PDI, Inc.:
1. Wipe the probe, cable, strain relief, and connector with a Super Sani-Cloth
®
Germicidal Disposable Wipe. Use
additional fresh wipes as needed.
2. Make sure the treated surface remains visibly wet for a minimum of two (2) minutes paying attention to seams,
gaps, gasket material, and recessed areas.
3. Use additional fresh wipes as needed to ensure continuous two (2) minutes of contact time.
4. Air dry the probe. Alternatively, use a soft non-linting cloth, blot the lens dry. Do not wipe the lens. Dry the rest
of the probe, cable, strain relief, and connector.
5. Once clean and disinfected, visually inspect the probe, strain relief, cable, and connector for signs of damage or
wear.
To disinfect the probe using the Intermediate-Level Disinfection (ILD) method with bleach (0.6% Sodium
Hypochlorite) and clean non-linting wipes:
1. Wipe the probe, cable, strain relief, and connector using a clean non-linting wipe wetted (damp but not dripping)
with bleach (0.6%). Use additional fresh wipes as needed.
2. Make sure the treated surface remains visibly wet for a minimum of ten (10) minutes paying attention to seams,
gaps, gasket material, and recessed areas.
3. Use additional fresh wipes as needed to ensure continuous ten (10) minutes of contact time.
4. Air dry the probe. Alternatively, use a soft non-linting cloth, blot the lens dry. Do not wipe the lens. Dry the rest
of the probe, cable, strain relief, and connector.
5. Once clean and disinfected, visually inspect the probe, strain relief, cable, and connector for signs of damage or
wear.
Cleaning and Disinfecting the Probe at the Point-of-Use
Maintenance 78