EasyManua.ls Logo

Butterfly Network iQ - Butterfly Iq Educational View Guidance

Butterfly Network iQ
137 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 9. Volume Measurement Accuracy
Volume Range Specification
0-100mL ±7.5 mL
101-740 mL ±7.5 %
The measurement accuracy specifications assume the tool is being used to scan a tissueequivalent phantom as instructed.
The volume range of the 3D Auto Bladder Volume Tool is 0-740ml. Although higher volumes may be estimated and displayed, Butterfly Network
cannot guarantee the accuracy of measurements outside of this stated range.
10.4. Butterfly iQ Educational View Guidance
CAUTION!
The Educational View Guidance tools are for Educational Use Only. Not intended for clinical or
diagnostic use.
NOTE
Educational View Guidance is not available in the United States.
The Educational View Guidance tools provide users with a visual indication of the quality of the image while
scanning with the Butterfly iQ/ iQ+/ iQ3. The Educational View Guidance tools support the following views:
Cardiac Apical 4 Chamber
Cardiac Parasternal Long Axis
Cardiac Parasternal Short Axis
Lung A-Lines/B-Lines
As you scan the subject, the tool provides real-time feedback on image quality using a red to green scale, with green
indicating a high-quality image . It indicates the proportion of experts who would assess the anatomical
view as measurable.
Accessing Educational View Guidance
Butterfly iQ Educational View Guidance
AI-Assisted Tools 65

Table of Contents